Networked infusion pumps start talking to an unknown host
The prompt — read this to the room
Your SOC flags outbound connections from 40 networked infusion pumps on the oncology and intensive care floors to an IP address in a hosting range nobody recognises. The traffic started after last week's vendor maintenance visit. As far as nursing staff can tell, the pumps are still delivering medication normally. The manufacturer's support line says a patch is 'in validation' and will not commit to a date. Clinical engineering owns the pumps, IT owns the network, and the vendor holds the admin passwords. Pulling the pumps means moving patients to manual pumps you do not have enough of. What do you do in the next hour?
Who to involve
Most important first.
- Incident response lead
- OT / plant engineer
- SOC analyst
- IT operations
- CISO / Head of security
- Vendor / procurement manager
What it tests
- Assess patient-safety risk and decide on device isolation together with clinical engineering, not around it
- Contain suspicious device traffic with network controls that keep the pumps working
- Hold the manufacturer to its contractual and regulatory duties for vulnerability response and evidence
- Preserve device and network evidence without breaking device validation or warranty
Outline of activities
- 1Alert triage: what the pumps are doing and since when
- 2Patient-safety risk call with clinical engineering and nursing
- 3Network containment for devices you cannot patch
- 4Vendor escalation, remote-access review and evidence
- 5Recovery, monitoring plan and the asset-inventory gaps it exposed
Free preview from the run sheet
First inject · minute 15
Flow data
Firewall export: 40 hosts in VLAN 212 (INF-PUMP) sending HTTPS to 203.0.113.48 every 300 seconds, about 2 KB out and 40 bytes back per session, first seen 6 days ago at 14:22, 18 minutes after the vendor engineer badged into the biomed workshop.
Questions for the Incident response lead
- What is your working hypothesis for the traffic, and what evidence would disprove it?
- How do you collect evidence from a device you cannot install tools on?
- What is your containment step that does not change how the pump delivers medication?
- When do you treat this as an incident across the hospital group rather than one site?
Built on: NIST SP 1800-8, Securing Wireless Infusion Pumps in Healthcare Delivery Organizations; FDA, Postmarket Management of Cybersecurity in Medical Devices (guidance); FDA, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (guidance); MDS2 form, Manufacturer Disclosure Statement for Medical Device Security (ANSI/NEMA HN 1-2019); HHS 405(d) HICP Technical Volume 2, medical device security practices; NIST SP 800-61r3.
The full run sheet below has every timed inject, the facilitator guide and questions for each role.
Run sheet
The facilitator's timeline of execution, the detailed steps for running the session and the questions to ask each person.
Loading…